Epithod®616 D-Dimer Kit is an in vitro diagnostic (IVD) test for the rapid and quantitative measurement of the fibrin degradation product D-Dimer in the plasma. 

Specifications

Simple Procedure

  Principle
 Immuno-chromatography
  Sample Type
 Serum/plasma, Whole blood (capillary and venous)
  Sample Volume
 5 𝛍L
  Test Result
 CRP (C-reactive protein)
  Test Range
 2.5 - 200 mg/L
  Precision
CV < 5%
  Test Time
Less than 2 minutes
  Calibration 

Olympus AU 5800, 

ELITECH Selectra ProM

Performance

Precision
Within-run
Day-to-day
Lv.1
Lv.2
Lv.3
Lv.1
Lv.2
Lv.3
CRP (mg/L)
2.84
46.83
113.70
2.94
46.38
112.76
SD (mg/L)
0.07
1.02
1.63
0.08
0.78
2.19
CV (%)
2.60
2.17
1.43
2.84
1.69
1.95

Contents

Name
Quantity
Package
Remark
TC/Test Cartridge
50 x 1
Individual pouch
Disposable
R1/Reagent
50 x 1
Zipper bag
Disposable
R2/Conjugates
1
Bottle
For 50 tests
R3/Wahing Solution
1
Bottle
For 50 tests
IC/Item Card
1
-
-
IFU/Instructions for use
1--

Test Principle